ASTM F2723

ASTM F2723

Tibial Baseplate/Insert Resistance to Dynamic Disassociation

Test method for evaluating the potential of mobile TKR bearing disassociation under repeated forces at high flexion. Five samples are tested for a total of 220,000 cycles, equivalent to 30 extreme motions per day for 20 years, or until failure of the retention mechanism.